In regards to gel crawling caused by the glass plates being stripped of
their ionic charge we have found that by placing the plate in acid wash
for at least 1 hour and then applying 'Coatasil' the problem has been
rectified. Plates are re-acid washed and coatasiled every 2 months (the
plates are used at least 4 times a week for Genescan runs). Hope this
solves your problems!