PROBLRMS WITH BIO-RAD SEQUI-GEN
Klaus.Matthaei at ANU.EDU.AU
Klaus.Matthaei at ANU.EDU.AU
Tue Feb 22 21:19:39 EST 1994
>We have had a similar experience. Ususally some acrylamide will
>accumulate in the top reservoir when we have new people using it. The
>best solution that we have found is to remove the accumulated
>acrylamide with a bent coat hanger and pray that the platinum wire
>doesn't get busted. So far so good. Has anyone else found a better
>solution?
>
>
Yep
Prevent the acrylamide from polymerising by pouring some water (70% ethanol
works even better) down the back before you start to pour the acrylamide
between the plates.
If you do get some polymerised a dilute solution of NaOH will eventually
dislodge it.
Cheers, Klaus
--------------------------------------------------------------------------
Klaus Matthaei
Gene Targeting
The John Curtin School of Medical Research
The Australian National University
E-mail: Klaus.Matthaei at anu.edu.au
"Think twice - Do once"
--------------------------------------------------------------------------