Biacore X sample injection
Andy Scotter
pcxajs at unix.ccc.nottingham.ac.uk
Tue Mar 13 03:22:16 EST 2001
I usually add 2-3ul of air then 2-3ul sample (additional to the
60ul in your example) then another 2-3ul of air when I inject samples into
a BIAcoreX. That is enough to get well defined bubbles in the tip either
side of a plug of your sample. That will prevent any mixing of your
sample with the buffer already in the system so you get better
sensorgrams.
Hope that helps,
Andy
********************************