Fluorescence microscopy question

Ian Clements ian at probes.com
Mon May 21 12:32:11 EST 2001

You can try a DAPI/Hoechst bandpass set (ex ~330-360, dm ~420, em
~430-480) or a DAPI long pass set (ex ~330-360, dm ~420, em >430).  These
wavelengths are approximate as different vendors have slightly different

Check out the following websites for more information.


oswaldo wrote:

> Hi there,
> Anybody knows which fluorescence filter set is the best to conduct
> observations on uvitex2B or calcofluor-stained slides for diagnosis of
> fungi/microsporidia?.
> Many thanks
> Oswaldo