	On the subject of getting the 'O' ring onto dialysis buttons, we have
used 10mm pyrex brand test tubes ( No. 9820) for years. Place the dialysis
membrane over the open end of the test tube. Roll the 'O' ring onto the test
tube, allowing the 'O' ring to hold the membrane. Firmly hold the membrane
covered tube onto the dialysis button, then roll the 'O' ring up the tube,
fold the membrane down and roll the 'O' ring down the tube and into its
groove in the button.  

Gerry Bunick                                       bunick at biovx1.bio.ornl.gov
Coordinator, Structural Biology Programs
Oak Ridge National Laboratory 