SAXS-Calibration

Victor Skita SKITA at BSAC.UCHC.EDU
Sat May 21 10:13:20 EST 1994


> Hello out there,
> I am looking for a calibration-standard (-sample) for
> SAXS-messurements. Any comment, any hint will be helpful!
> 
> Thanks in advance
> 
> Wolfgang
> 
> -- 
> +--------------------------------------------+
> |  Wolfgang Sch"utze   wo.schuetze at tu-bs.de  |                      
> |  Pharmaceutical Technology                 |
> |  University Braunschweig/Germany           |
> +--------------------------------------------+


Can I have more information? How far out in recprical space do you plan
to measure (ie. how small is the small in small angle)? For example, mica
(from NIST) diffracts beautifully at 1/10 A-1 and is "rock" solid. Please
send details. (Folks often use lead-sterate, but it must be DRY. The "d"
spacing will change with hydration.)

Vic.

===============================================================================