>We have tried but couldn't detect DsRed by flow cytometry.
>DsRed probably isn't excited enough at 488 nm.
>>Gregers J. Gram
>Laboratory for Infectious Diseases 144
>H:S Hvidovre Hospital, Denmark
>www.inflab.suite.dk/
>------------------------------
Excitation efficiency of DsRed should still be 30-40% at 488nm. So if
you can't see anything at that wavelength, it means that you would at
best get a very bad S/N when exciting at the optimal wavelength. Is
your emission filter (and dichroic mirror, in case there is one in
the system) suitable for DsRed?
Beat
LIFE IMAGING SERVICES
- visit our web site at http://www.lis.ch
+-----------------------------------------------------------
| Dr. B. Ludin
| Life Imaging Services fon ++41 (0)79 235 7154