[Neuroscience] Re: Ideas for getting a Sutter to pull reasonable
pipettes?
Richard Vickery
via neur-sci%40net.bio.net
(by Richard.Vickery from unsw.edu.au)
Tue Jul 3 07:04:57 EST 2007
Have you tried a two stage pull? Pull it but not so it breaks, then pull
again. On the last Flaming Brown I used it could pull with a different set of
parameters on each stage.
Cheers
Richard
Bill wrote:
> Hey,
>
> I've got a sutter P-97, Flaming/Brown, Puller. I'm using thin walled
> 1.5mm OD borosilicate glass (hHrvard glass). I've tried every
> combination of velocity, temperature and pull I can think of, and most
> of the ones in the"recipe book". The problem is the tips are too big,
> or to get the tips to a reasonable size, the tapper is so long then
> walls of the electrode are parallel by the end, and the resistance is
> massive.
>
> I know no one can give me exact numbers, but does anyone have any
> general ideas?
>
> Anything appriciated,
>
> -Bill
>
More information about the Neur-sci