>Dear bionetters,
>
>This is not a yeast question, per se, but I am sequencing a yeast gene. Anyway,
>when I sequence, I routinely load one set of sample, and load a second set
>after the xylene cyanol marker of the first set has run out of the gel. When I
>am ready to load the second set, however, the top of the gel (the part that the
>comb touches to form the wells) has become distorted. I do not know if this is
>due to the heat or the electrical current, but it interferes with loading of
>the second set of samples.
>
>I've found that running the gel on a lower wattage helps, though not
>completely. I am using a wedge spacer on a 6% gel, and running at 50 or less
>watts (about 1200V and 40mA).
>
>Thanks in advance for any and all help.
>
>KC.
>
>P.S. Another problem I have is that the gel also developes air bubbles several
>centimeters from the top after extended runs. Any remedies?
>
I had had this kind of problem. But it seemed localised in the particular gel
because others do not had it. You can forgive me if I say to you that I
run my gels at 63W (1900-2300V, near 50 degrees) and use not to have
problems. However, try to change acrylamide or Tris.
Manuel
-------------------------------------